Transmitter for B2H6 Diborane Gas Detection. Semiconductor
gas monitoring applications. Long life electrochemical
sensor. Now available suitable for use in
|
no false
alarms electrochemical B2H6 Diborane gas sensor technology
|
|
Self-Calibration adjusts span monthly based on sensor life
curve
|
|
Auto-Gas Calibration, non-intrusive, hands-free, magnetic
switches
|
|
Not affected by temperature -20 deg C to +40 deg C, stable by
design
|
|
LCD display - 12 characters x 2 lines. Provides user
interface with magnetic switches - no dip switches
|
|
Offsite sensor calibration with memory chip embedded in
sensor
|
|
Peak value, 15-min. TWA, Remaining Sensor Life, Replace
Sensor indication and number days since last gas calibration
|
|
Optional Relay Module in same enclosure with low, mid, high
and fault relays
|
|
Can be used with Mil-Ram's Wireless Telemetry System
|
|
Available suitable for use in SIL 2
environments
|
|
Continuous advanced Diagnostics
|
|
Detection Range
|
0.00-1.00 PPM Diborane Gas
|
|
Resolution
|
0.01 PPM Diborane
|
|
Response Time
|
< 30 sec to 90% of final reading
|
|
Recovery Time
|
< 30 sec. to 90% recovery
|
|
Temperature Range
|
-20 deg.C to +40 deg.C
|
|
Humidity Range
|
15-95% RH, non-condensing, continuous
|
|
Power Requirement
|
12V or 15-30VDC; 1.08W @24VDC
|
|
Output
|
4-20mA linear. Optional Modbus RTU. Optional Relays; low, mid, high & fault.
|
|
Long life
|
>2 years typical under normal operating